Pacemaker Therapy

|
|
|
|
The most common indication for pacing in children is
complete heart block although the
sick sinus syndrome following atrial switch procedures (Senning
and Mustard operations) is still relatively common. |
|
When permanent pacing is required, the usual approach is
via the left sublcavian vein. A pacing wire is fed into the ventricle and
connected to a programmable generator (“pacing box”) implanted under the
skin. The generators are now very small (approximately 50 mm
in diameter and 8 mm thick. In dual chamber pacing an additional atrial
lead is used. A pacing check is performed at implantation, one day later
and then at regular intervals thereafter.
|

|
|
The generator can be programmed for different rates,
voltage output, mode, sensitivity and rate responsiveness using a
transmitter placed close to the skin. It can also be interrogated for
patient events. The life of a generator is variable and depends on its
output and frequency of use but is usually 5-10 years. |
|
If small infants require pacing then the morbidity from a
tranvenous approach may be considerable. The surgeon can then place a lead
on the ventricle (epicardium) and the generator in the abdomen. These can
readily be removed when the child is older and a transvenous system
implanted.

|

|
|
|
|
Pacing modes follow an internationally agreed convention.
Four letters are used. The first letter refers to the paced chamber, the second to the chamber where
intrinsic activity is detected (sensing). The third refers to how the pacemaker reacts to the
heart’s intrinsic electrical activity – usually inhibited. The fourth
letter is "R" and denotes rate responsiveness to exercise. The common
modes are VVI (paces & senses ventricle, inhibited by ventricular
contraction), VVIR and DDD (paces & senses atrium & ventricle, inhibited by atrial
&
ventricular contraction). This is the most physiological mode. AAI
mode may be used in the sick sinus syndrome provided the AV node is
normal. |

 |
|
Pacing is reflected on the ECG with a spike preceding the
paced chamber. The above example shows a pacing spike initiating
ventricular contraction (VVI mode in a patient with complete heart block). |
|
Pacemakers may become infected. This is most common around
the time of implantation. It presents with fever, local inflammation, pain
swelling and tenderness if the infection is sited around the pacemaker box
or lead insertion site. It may also present as infective endocarditis with
vegetations on the pacing lead (arrow). The system may need to be
explanted to enable antibiotics to control the infection.
|
 |
|
Failure to pace successfully may present with syncope or
presyncope or be asymptomatic and noted at the pacemaker check or a
routine ambulatory ECG. It may be due to generator failure (e.g. low
battery), exit block (fibrosis between lead and myocardium), lead
displacement or fracture (arrow). No pacing spikes are seen on the ECG.
Inappropriate sensing may also cause difficulties. If oversensing occurs
the pacemaker may fail to pace when required. If undersensing the
pacemaker may fail to sense at all pace when not required.
|

 |
|
Temporary pacing is most often required in the immediate
postoperative period. It is such a common problem that temporary leads are
placed on the atrium and ventricle and passed out through the chest wall
at the time of surgery. They are removed by traction when no longer
needed. Outside of the immediate post-operative period temporary pacing is
rarely indicated. If necessary a lead (arrow) is inserted via the
subclavian, jugular or femoral vein and connected to an external pacemaker
box.
|
 |
|
|
|
 |
|
This page was
last edited
14/2/2004 |